CINE STANDARD
- Introduction
- Product Software
CINE STANDARD
REFLECTANCE
The test patterns TE142 is supplied on sheets of paper or cardboard and backed with
self-adhesive foils if requested. The individual elements are used for assembling test
boards or test walls according to specific ideas and requirements. The self-adhesive
tape facilitates an easy mounting of the test pattern by hand.
For evaluating the resolution power of the film it is necessary to define the enlargement
factor at first. For the size of the test pattern TE142 the recommended factor is 100x or
150x. Based on the film format (e.g. S16: 12.3 x 740 mm). The test pattern then have to
be fixed within these limits, possible in combination with other test patterns (e.g. sector stars).
The test pattern shows 9 vertical and 9 horizontal groups of line pairs, marked C to K.
The gradation of these line pairs follows the factor √2, i.e. the number of lines/mm doubles
every two steps. The following values apply to the line pairs when using the reduction
factors indicated:
|
|
Resolving power in line pairs/mm with factor |
|
|
|||
field |
original |
|
50x |
100x |
150x |
200x |
|
C |
1.4 |
|
70 |
140 |
210 |
280 |
|
D |
1.0 |
|
50 |
100 |
150 |
200 |
|
E |
0.7 |
|
35 |
70 |
105 |
140 |
|
F |
0.5 |
|
25 |
50 |
75 |
100 |
|
G |
0.35 |
|
17.5 |
35 |
52.5 |
70 |
|
H |
0.25 |
|
12.5 |
25 |
37.5 |
50 |
|
I |
0.175 |
|
8.75 |
17.5 |
26.25 |
35 |
|
J |
0.125 |
|
6.25 |
12.5 |
18.75 |
25 |
|
K |
0.0875 |
|
4.375 |
8.75 |
13.125 |
17.5 |
|
|
|
|
|
|
|
||
TE142/2 |
|
Resolving power in line pairs/mm with factor |
|
|
|||
field |
original |
|
50x |
100x |
150x |
200x |
|
C |
1.4 |
|
70 |
140 |
210 |
280 |
|
D |
1.0 |
|
50 |
100 |
150 |
200 |
|
E |
0.7 |
|
35 |
70 |
105 |
140 |
|
F |
0.5 |
|
25 |
50 |
75 |
100 |
|
G |
0.35 |
|
17.5 |
35 |
52.5 |
70 |
|
H |
0.25 |
|
12.5 |
25 |
37.5 |
50 |
|
I |
0.175 |
|
8.75 |
17.5 |
26.25 |
35 |
|
J |
0.125 |
|
6.25 |
12.5 |
18.75 |
25 |
|
K |
0.0875 |
|
4.375 |
8.75 |
13.125 |
17.5 |
|
TE142/3Resolving power in line pairs/mm with factor
field |
original |
50x |
100x |
150x |
200x |
C |
2.0 |
100 |
200 |
300 |
400 |
D |
1.4 |
70 |
140 |
210 |
280 |
E |
1.0 |
50 |
100 |
150 |
200 |
F |
0.7 |
35 |
70 |
105 |
140 |
G |
0.5 |
25 |
50 |
75 |
100 |
H |
0.35 |
17.5 |
35 |
52.5 |
70 |
I |
0.25 |
12.5 |
25 |
37.5 |
50 |
J |
0.175 |
8.75 |
17.5 |
26.25 |
35 |
K |
0.125 |
6.25 |
12.5 |
18.75 |
25 |
- Previous:CINE RESOLUTION TEST CHART
Next:ISO TEST PATTERN
News
- 2017-03-31 sales elite awards n the first quar ...
- 2014-06-26 3nh TUV Certificate
- 2014-06-12 Colorimeters Versus Spectrophotomet ...
- 2014-06-26 3nh Products Have Got CE Certificat ...
- 2014-06-30 NS800 Spectrophotometer CE Certific ...
- 2014-06-12 Human Eyes to Distinguish Colors
- 2015-05-08 3nh Products FCC Certificate
- 2015-05-08 PANTONE CU Color Card (Electronic E ...